Figure 4-11. Anatomy of a measurement
Power Line Rejection Command/Response mode
min. Read Interval(s) / max
meas. rate(Hz)
H/W Trigger mode min. Read
Interval(s) / max meas. Rate
(Hz)
Aperture 60
Hz
50Hz 400Hz
5.1200s [1]
√ √ √
5.121s / 0.2 N/A
5.0666s [1]
√
5.0677s / 0.2 N/A
2.08s [1]
√ √
2.081s / 0.5 N/A
2.0s [1]
√ √ √
2.001s / 0.5 N/A
1.06666s [1]
√
1.067s / 1 N/A
960ms [1]
√ √
0.9605s / 1 N/A
533.33ms [1]
√
533.6ms / 2 N/A
480ms [1]
√ √
480.2ms / 2 N/A
266.666ms [1]
√
268ms / 4 N/A
160.0ms
√ √ √
166ms / 6 160.3 ms / 6
133.33ms
√
134ms / 8 133.5 ms / 8
80.00ms
√ √
80.4ms / 13 80.2 ms / 13
66.6667ms
√
67.2ms / 15 66.713 ms / 15
40.00ms
√ √
40.4ms / 25 40.32 ms / 24.8
33.333ms
√
33.643ms / 29.72 33.38 ms / 30
20.00ms
√ √
20.098ms / 49.76 20.33 ms / 50
16.6667ms
√
16.77ms / 59.6 16.89 ms / 59
10ms 10.094ms / 99 10.25 ms / 97
8.333ms 8.422ms / 119 8.503 ms / 115
5ms 5.109ms / 195 5.187 ms / 185
4.16667ms 4.265ms / 234 4.274 ms / 220
2.5ms 2.598ms / 385 2.614 ms / 350
2.0833ms 2.177ms / 458 2.216 ms / 410
1.25ms 1.344ms / 744 1.380 ms / 625
1.0417ms 1.133ms / 880 1.158 ms / 864
625µS
719µs / 1,390 728 µs / 1,370
520.83µS
617µs / 1,625 622 µs / 1,610
312.5µS
410µs / 2,445 414 µs / 2,445
260.42µS
355µs / 2,825 358 µs / 2,825
130.21µS
215µs / 4,660 217 µs / 4,660
2.5µS
47µs / 21,600 45 µs / 22,200
[1] Not available with any of the Triggered modes.
Table 4.1: The SM2064 has 31 A/D apertures available, ranging from 5 Seconds to 2.5uSec