Operator`s manual

31 Signametrics
“Specifications” for details.). When measuring RMS values, there is no point setting the Read Interval (1/rate) to a
value shorter than 0.16s since the RMS circuitry has a settling time that is greater.
Range: Can be set to Autorange or manual by clicking on the appropriate range in the lower part of the Windows
panel. Autoranging is best used for bench top application and is not recommended for an automated test
application due to the uncertainty of the DMM range, as well as the extra time for range changes. Locking a range
is highly recommended when operating in an automated test system, especially to speed up measurements. Another
reason to lock a range is to control the input impedance in DCV. The 240 mV and 2.4 V ranges have virtually
infinite input impedance, while the 24 V and 240 V and 330 V ranges have 10 M input impedance.
S_Cal: This function is the System Calibration that corrects for internal gain, scale factor and zero errors. The
DMM does this by alternatively selecting its lo l DC reference and a zero input. It is required at least once every
ay to meet the SM2060 accuracy specifications. It is recommended that you also perform this function whenever
e external environment changes (e.g. the temperature in your work environment changes by more than 5°C, or the
M2064 on board temperature sensor indicates more than a 5°C change). This function takes less than a few
conds to perform. Disconnect all leads to the DMM before doing this operation. Keep in mind that this is not a
bstitute for periodic calibration, which must be performed with external standards.
losedLoop: This check box selection is used in conjunction with the AC and DC Voltage-Source functions of the
M2064. When checked, the DMM monitors the output level and continuously applies corrections to the output
vel. When not checked, the DMM is a 12-bit source vs. 16 bits in the ClosedLoop mode.
penCal: This check box selection is used in conjunction with inductance measurement. It is necessary to perform
pen Terminal Calibration using this control, prior to measuring inductance. This function characterizes both the
ternal DMM circuitry as well as the probe cables. To perform OpenCal, attach the probe cables to the DMM,
enCal button.
ca
d
th
S
se
su
C
S
le
O
O
in
leaving the other end of the probe cables open circuited. Then, activate the Op
Sources Panel:
There are three function buttons in the Source group (SM2064 only). The V, I, LEAK buttons
select one of three source functions, Voltage (DC and AC), IDC and Leakage. The Sources Panel is automatically
enabled when one of the source functions is enabled. It can also be invoked using the Sources Panel selection
under the
Tools menu. This panel allows the entry of values for all of the source functions, including Leakage.
The V-OUT Scroll bar and Text box are used to
set the Voltage for DC and AC Volts as well as
for Leakage. When sourcing ACV, the voltage is
used to control inductance frequency. When
, use the I-OUT set of
hen measuring timing or freqeuncy
in RMS and the FREQ. Scroll bar and Text box
control the frequency of the source. It is also
sourcing DC current
controls. W
the THRESH set of controls is used for
comperator threshold. All of the source controls
are context sensitive and will be enabled when