Operator`s manual

The SM2060 has are 26 A/D apertures available, ranging from 5 Seconds to 625uSec. The following
table contains all available measurement apertures corresponding minimum read intervals and
measurement rates.
Power Line Rejection Command/Response
mode min. Read
Interval(s) / max meas.
rate(Hz)
H/W Trigger mode min. Read
Interval(s) / max meas. Rate
(Hz)
Aperture 60Hz 50Hz 400Hz
5.1200s [1]
5.121s / 0.2 N/A
5.0666s [1]
5.0677s / 0.2 N/A
2.08s [1]
2.081s / 0.5 N/A
2.0s [1]
2.001s / 0.5 N/A
1.06666s [1]
1.067s / 1 N/A
960ms [1]
0.9605s / 1 N/A
533.33ms [1]
533.6ms / 2 N/A
480ms [1]
480.2ms / 2 N/A
266.666ms [1]
268ms / 4 N/A
160.0ms
166ms / 6 160.3 ms / 6
133.33ms
134ms / 8 133.5 ms / 8
80.00ms
80.4ms / 13 80.2 ms / 13
66.6667ms
67.2ms / 15 66.713 ms / 15
40.00ms
40.4ms / 25 40.32 ms / 24.8
33.333ms
33.7ms / 30 33.38 ms / 30
20.00ms
20.35ms / 50 20.33 ms / 50
16.6667ms
16.9ms / 59 16.89 ms / 59
10ms 10.36ms / 97 10.25 ms / 97
8.333ms 8.68ms / 115 8.503 ms / 115
5ms 5.36ms / 185 5.187 ms / 185
4.16667ms 4.52ms / 220 4.274 ms / 220
2.5ms 2.86ms / 350 2.614 ms / 350
2.0833ms 2.44ms / 410 2.216 ms / 410
1.25ms 1.6ms / 625 1.380 ms / 625
1.0417ms 1.39ms / 719 1.158 ms / 864
625µS
917µs / 1,090 728 µs / 1,370
[1] Not available with any of the Triggered modes.
Precise control of the measurement timing and line frequency rejection can be accomplished by controlling the Read
Interval and Aperture. Line rejection is dictated by the Aperture, and the duration of measurement is controlled with
Read Interval
.
Read Interval can be programmed in
µs increments for values up to 65ms
, and in 20
µ
s increments to 1 second.
Figure 2-1: Time frame of a single measurement.