2.10.2 Frequency and Period Measurement
ACV Mode
Input Impedance 1 M with < 300 pF
Frequency Range 2 Hz - 100 Hz 100 Hz-1 kHz 1 kHz-10 kHz 10 kHz-100 kHz 100 kHz-300 kHz
Resolution 1 mHz 10 mHz 100 mHz 1 Hz 1 Hz
Uncertainty is ±0.002% of
reading ± adder shown
4 mHz 20 mHz 200 mHz 2 Hz 5 Hz
Input Signal Range [1] 10% - 200%
of range
10% - 200%
of range
10% -200%
of range
10% - 200%
of range
45% -200%
of range
[
1] Input RMS voltage required for a valid reading. Do not exceed 250 V RMS input. For example, 10% -200%
of range indicates that in the 240 mVAC range, the input voltage should be 24 mV to 660 mV RMS.
ACI Mode
Input Impedance 10 in the 3 mA and 30 mA ranges, 0.1 in the 330 mA and 2.5 A ranges.
Frequency Range 2 Hz - 100 Hz 100 Hz-1 kHz 1 kHz-10 kHz 10 kHz-500 kHz
Resolution 1 mHz 10 mHz 100 mHz 1 Hz
Uncertainty 0.01% ±4 mHz 0.01% ±20 mHz 0.01% ±200 mHz 0.01% ±2 Hz
Input Signal Range,
2.4 mA, 240mA Ranges
[1]
10% -500%
of range
10% - 500%
of range
10% -500%
of range
10% - 500%
of range
Input Signal Range,
24 mA, 2.4 A ranges
50% -100%
of range
50% - 100%
of range
50% - 100%
of range
50% - 100%
of range
[1] Input current required to give a valid reading. For example, 10% -500% of range indicates that in the 3.3 mA
range, the input current should be 0.33 mA to 16.5 mA.
2.10.3 Duty Cycle Measurement
Frequency Range 2 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz to 10 kHz 10 kHz to 100 kHz
Resolution 0.02% 0.2% 2% 20%
Typical Uncertainty is
±0.03% of reading ±
adder shown
0.03% 0.3% 3% 20%
Full scale reading 100.00 % 100.00 % 100.00 % 100.00 %
2.10.4 Pulse Width
± (% of reading + sec)
Polarity Frequency range Resolution Width range Typical
Uncertainty
Positive or negative pulse
widths
2 Hz to 100 kHz
1 µs 2 µs to 1 s 0.01 +/- 4 µs