Operator`s manual

17 Signametrics
Ro: Variable from 10
to 10 k
Measurement Method: 4-Wire
Temperature units: Selectable
o
C or
o
F
RTD Type
Ro ()
Resolution Temperature
range
Temperature Accuracy 23°C ± 5°C [1]
One Year
pt385, pt3911,
pt3916, pt3926
100, 200 0.01°C -150 to 650°C ±0.06°C
pt385, pt3911,
pt3916, pt3926
500, 1 k 0.01°C -150 to 650°C ±0.03°C
Cu (Copper)
Less than 12 0.01°C -100 to 200°C ±0.18°C for temperatures 20°C, ±0.05°C
otherwise
Cu (Copper)
Higher than 90 0.01°C -100 to 200°C ±0.10°C for temperatures 20°C, ±0.05°C
otherwise
[1] With Aperture of 0.5s and higher, using a 4-wire RTD. Measurement accuracy does not include
RTD probe error.
2.8 Thermocouple Temperature Measurement (SM2064)
Cold Junction Compensation: By Sensor measurement or S/W setting.
Cold Junction Temperature range: 0
o
C
to 50
o
C
Cold Junction Sensor: Use SMX40T or SM40T Isothermal unit, or define sensor equation
Isothermal Block compatibility: SM4022, SM4042, SMX4032, SM40T, SMX40T
Temperature units: Selectable
o
C or
o
F
TC Type Resolution Maximum Temperature
[2]
Temperature Accuracy 23°C ± 5°C [1]
One Year
B
0.01°C 2200°C ±0.38 °C
E
0.01°C 1200°C ±0.035 °C
J
0.01°C 2000°C ±0.06 °C
K
0.01°C 3000°C ±0.07 °C
N
0.01°C 3000°C ±0.10 °C
R
0.01°C 2700°C ±0.25 °C
S
0.01°C 3500°C ±0.35 °C
T
0.01°C 550°C ±0.06 °C
[1] With Aperture of 0.5s and higher. Measurement accuracy does not include Thermocouple error.
[2] DMM Linearization temperature range may be greater than that of the Thermocouple device.
2.9 Additional Component Measurement Capability
2.9.1 Diode Characterization
Available DC current values 100 ηA, 1 µA, 10 µA, 100 µA and 1 mA.
SM2064 add variable current of 10 ηA to 12.5 mA
Typical Current Value Uncertainty 1%
Typical Voltage Value Uncertainty 0.02%
Maximum diode voltage compliance 4 V
2.9.2 Capacitance, Ramp Method (SM2064)
Accuracy ± (% of reading + Farads) [1]