Operator`s manual
151 Signametrics
The SM60CAL.DAT file is created by performing external calibration. The general calibration
algorithm consists of applying a zero value to the DMM followed by a value of 2/3
P
rd
P
of the top of each
range. Calibration of your SM2060/64 is best performed using calibration software available from
Signametrics.
When using multiple DMMs in a single chassis, the SM60CAL.DAT file must have a calibration record
for each DMM. You can combine the unique calibration records of each DMM into one
SM60CAL.DAT file using any ASCII text editor such as “notepad.exe”.