Operator`s manual
13 Signametrics
AC Volts Accuracy with Fast RMS disabled (default).
With Fast RMS disabled, settling time to rated accuracy is within 0.5 s.
Accuracy ± (% of reading + Volts) [1]
Range Frequency 24 hours
23°C ± 1°C
90 Days
23°C ± 5°C
One Year
23°C ± 5°C
10 Hz - 20 Hz
3.0 + 350 µV 3.1 + 380 µV 3.2 + 430 µV
20 Hz - 47 Hz
0.92 + 150 µV 0.93 + 170 µV 0.95 + 200 µV
47 Hz - 10 kHz
0.13 + 100 µV 0.14 + 110 µV 0.15 + 120 µV
10 kHz - 50 kHz
0.55 + 160 µV 0.6 + 200 µV 0.63 + 230 µV
240 mV
50 kHz - 100 kHz
5.3 + 350 µV 5.4 + 370 µV 5.6 + 400 µV
10 Hz - 20 Hz 3.0 + 2 mV 3.1 + 2.2 mV 3.2 + 2.5 mV
20 Hz - 47 Hz 0.93 + 1.3 mV 0.96 + 1.5 mV 1.0 + 1.7 mV
47 Hz - 10 kHz 0.05 + 1 mV 0.055 + 1.1 mV 0.065 + 1.2 mV
10 kHz - 50 kHz 0.62 + 1.2 mV 0.65 + 1.3 mV 0.70 + 1.5 mV
2.4 V
50 kHz - 100 kHz 5.1 + 1.5 mV 5.2 + 1.7 mV 5.3 + 2 mV
10 Hz - 20 Hz 3.0 + 14 mV 3.1 + 16 mV 3.3 + 20 mV
20 Hz - 47 Hz 0.93 + 12 mV 0.96 + 14 mV 1.0 + 16 mV
47 Hz - 10 kHz 0.06 + 10 mV 0.065 + 11 mV 0.073 + 13 mV
10 kHz - 50 kHz 0.31 + 18 mV 0.33 + 21 mV 0.35 + 25 mV
24 V
50 kHz - 100 kHz 2.0 + 30 mV 2.2 + 35 mV 2.4 + 40 mV
10 Hz - 20 Hz 3.0 + 140 mV 3.1 + 160 mV 3.3 + 200 mV
20 Hz - 47 Hz 0.93 + 120 mV 0.96 + 130 mV 1.0 + 150 mV
47 Hz - 10 kHz 0.04 + 100 mV 0.045 + 110 mV 0.06 + 130 mV
10 kHz - 50 kHz 0.32 + 150 mV 0.4 + 170 mV 0.45 + 200 mV
240 V
50 kHz - 100 kHz 2.5 + 200 mV 2.8 + 240 mV 3.2 + 300 mV
10 Hz - 20 Hz 3.0 + 200 mV 3.1 + 160 mV 3.3 + 200 mV
20 Hz - 47 Hz 1.0 + 180 mV 1.1 + 200 mV 1.1 + 250 mV
47 Hz - 10 kHz 0.05 + 150 mV 0.07 + 200 mV 0.08 + 230 mV
10 kHz - 50 kHz 0.34 + 200 mV 0.45 + 250 mV 0.5 + 300 mV
330 V
50 kHz - 100 kHz 2.5 + 270 mV 2.8 + 350 mV 3.2 + 400 mV
ACV Noise Rejection Common Mode rejection, for 50 Hz or 60 Hz with 1 kΩ imbalance in either lead, is better
than 60 dB.
AC Volts Accuracy with Fast RMS enabled.