2.3 Resistance Measurements
2.3.1 2-wire
Accuracy ± (% of reading + ) [1]
Range [4] Full Scale
7 ½ Digits
Resolution Source
current
24 hours
23°C ± 1°C
90 Days
23°C ± 5°C
One Year
23°C ± 5°C
24 [3] 24.000000 1 µΩ
10 mA
0.0038 + 1.4 m [2] 0.005 + 1.6 m [2] 0.008 + 2 m [2]
240 240.00000 10 µΩ
1 mA
0.0037 + 4.5 m [2] 0.0046 + 5 m [2] 0.007 + 6 m [2]
2.4 k 2.4000000 k 100 µΩ
1 mA
0.0023 + 28 m 0.004 + 32 m 0.006 + 33 m
24 k 24.000000 k 1 m 100 µA 0.0025 + 300 m 0.004 + 330 m 0.006 + 350 m
240 k 240.00000 k 10 m 10 µA 0.0055 + 3.2 0.006 + 4 0.007 + 5
2.4 M 2.4000000 M 100 m 1 µA 0.018 + 40 0.03 + 50 0.04 + 70
24 M 24.0000 M 100
100 nA
0.12 + 400 0.13 + 500 0.2 + 600
240 M[3] 240.000 M 1 k
10 nA
0.8 + 20 k 1.0 + 30 k 1.3 + 50 k
[1] With Aperture set to 0.5 Sec, and within one hour from Self Calibration (S-Cal).
[2] Use of S-Cal and Relative to improve measurement floor.
[3] Ranges are only available with the SM2064.
[4] Test voltages are 2.4V max with the exception of the 24
and 240
ranges 240 mV.
2.3.2 4-wire
Accuracy ± (% of reading + ) [1]
Range [4] Full Scale
7 ½ Digits
Resolution Source
current
24 hours
23°C ± 1°C
90 Days
23°C ± 5°C
One Year
23°C ± 5°C
24 [3] 24.000000 1 µΩ
10 mA
0.0038 + 0.7 m [2] 0.005 + 0.8 m [2] 0.008 + 1 m [2]
240 240.00000 10 µΩ
1 mA
0.0037 + 3 m [2] 0.0046 + 4 m [2] 0.007 + 5 m [2]
2.4 k 2.4000000 k 100 µΩ
1 mA
0.0023 + 28 m 0.004 + 32 m 0.006 + 33 m
24 k 24.000000 k 1 m 100 µA 0.0025 + 300 m 0.004 + 330 m 0.006 + 350 m
240 k 240.00000 k 10 m 10 µA 0.0055 + 3.2 0.007 + 4 0.007 + 5
2.4 M 2.4000000 M 100 m 1 µA 0.018 + 40 0.03 + 50 0.04 + 70
24 M 24.0000 M 100
100 nA
0.12 + 400 0.13 + 500 0.2 + 600
[1] With Aperture set to 0.5 Sec, and within one hour from Self Calibration (S-Cal).
[2] Use of Relative to facilitate indicated floor (adder part of spec).
[3] 24 range only available with SM2064.
[4] Test voltages are 2.4V max with the exception of the 24
and 240
ranges 240 mV.
2.3.3 6-wire Guarded Resistance Measurement (SM2064)
This is an in-circuit forced guard measurement method, as implemented in ICT testers. Add this typical additional
error to the above specification.
Accuracy ± (% of reading + )
Range Guard forced current
One Year 23°C ± 5°C [1]
24 [3]
10 mA
0.3 + 4 m
240
1 mA
0.003 + 20 m
2.4 k
1 mA
0.005 + 100 m
24 k 100 µA 0.03 + 1
240 k 10 µA 0.35 + 10
24 M 1 µA 0.85 + 1000
[1] This table should be used in conjunction with the 2-wire and 4-wire table above.