Operator`s manual

Signametrics 12
2.9 Other Specifications
Temperature Coefficient Less than 0.1 x accuracy specification below 18C and over
28C
Hardware Interface Single USB/cPCI 3U slot
Overload Protection (voltage inputs) 250 VDC, 250 VAC
Isolation 300 VDC, 250 VAC from Earth Ground
Maximum Input (Volt x Hertz) 8x10
6
Volt x Hz normal mode input (across Voltage HI &
LO).
1x10
6
Volt x Hz Common Mode input (from Voltage HI or
LO relative to Earth Ground).
Safety Designed to IEC 1010-1, Installation Category II.
Calibration Calibrations are performed by Signametrics in a computer at
23C internal temperature rise. All calibration constants are
stored in an onboard memory and in a text file.
Temperature Range Operating -10C to 65C
Temperature Range Storage -40C to 85C
Size Single 3U USB or CompactPCI slot
Power +5 volts, 200 mA maximum
Note: Signametrics reserves the right to make changes in materials, specifications, product functionality, or
accessories without notice.