User Manual
SIGLENT
36 SVA1000X Service Manual
Removing the Main Board, Channel Board
Figure 4-4 View of front panel subassembly with main board
Removal steps:
1. Place the analyzer back side down on a soft surface such as an anti-static
mat.
2. Remove the screws located on the metal shelf and each board.
3. Remove the RF board, TG board, Digital board from the metal shelf.