Tracking Generator (TG) Test
Specification
Frequency range
5 MHz ~1.5 GHz
Output level
-20 dBm ~ 0 dBm
Output flatness
+/-3 dB
SVA1000X
RF IN
TG OUT
N-N
Figure 3-10 Connecting test instruments for TG Output
Steps:
1. Connect the RF IN port and TG OUT port by an N-N cable as figure 3-10
shows
2. Configure the spectrum analyzer:
(a) Open the TG mode
(b) Set the span to full span
(c) Set the Ref level to 0 dB
(d) Set the RBW and VBW to Auto
3. Press Peak to find the maximum value of the TG output, record the
maximum value P1, check if P1 < TG level + 3dB
4. Press Peak -> Search config -> Peak type ->min, to find the minimum value
of the TG output, record the maximum value P2, check if P2 >TG level -
3dB
Record:
TG level
P1
P1-TG level
Pass/Fail
0 dBm
-10 dBm
-20 dBm
TG level
P2
P2-TG level
Pass/Fail
0 dBm
-10 dBm
-20 dBm