4. Configure the spectrum analyzer:
a) Set the center frequency to 50 MHz.
b) Set the span to 10 kHz.
d) Set the reference level to -30 dBm.
e) Set the input attenuation to 0 dB.
f) Set the resolution bandwidth to 300 Hz.
g) Set the video bandwidth to 10 Hz.
h) Set the sweep time to auto and the auto sweep time to accuracy.
5. Press Peak to find the maximum value P1. Then, Set the center frequency
to 100 MHz, Press Peak to find the maximum value P2.
Record:
P1
P2
P1 – P2