User Manual
SIGLENT
18 SVA1000X Service Manual
ERROR = A2 – P2
8. Frequency response = |GLOBAL ERROR - SYSTEM ERROR|, compare
the calculated result with the specification
9. Enable the preamplifier and set the output amplitude of the signal
generator to -40dBm. Repeat steps 3 to 8 and record the results.
Record:
Preamp off
Frequency
P1 (PM)
P2 (SA)
50 MHz
Frequency
A1(PM)
A2(SA)
A1 – P1
(System
Error)
A2 – P2
(Global
Error)
Frequency
Response
Pass/Fail
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
1.5 GHz
Preamp on
Frequency
P1 (PM)
P2 (SA)
50 MHz
Frequency
A1(PM)
A2(SA)
A1 – P1
(System
Error)
A2 – P2
(Global
Error)
Frequency
Response
Pass/Fail
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
1.5 GHz