10 SVA1000X Service Manual
USB Device Test
To test if the USB Device interface works normally.
Tools
A computer with USB interface that is compatible with running National
Instruments NI-MAX software
A standard USB cable (Type A-B)
NI-MAX software
Steps
1. Download and install National Instruments NI MAX software by following the
installation instructions provided by National Instruments.
2. Connect the analyzer USB Device port and the computer using an USB
cable.
3. Run NI MAX software. Click Device and interface at the upper left corner of
the NI software interface and immediately displays the USBTMC device
symbol.
4. Click “Open VISA Test Panel” option button, and then the following
Interface will appear. Then click the Input/Output option button and click
the “Query option button in order to view the Read operation information.