Programming Guide

Table Of Contents
SIGLENT
SVA1000X Programming Guide 59
span.
Gets dBc value.
Parameter
Type
Float
Parameter
Range
0.1~100
Return
Float
Default
26
Menu
Meas > Occupied BW > Meas Setup > dBc
Example
:OBWidth:XDB 3
3.7.3.4 Query OBW and Centroid (:MEASure:OBWidth?)
Command
Format
:MEASure:OBWidth?
Instruction
Use this command to query the occupied bandwidth and bandwidth centroid
according to the method you set.
Parameter
Type
None
Parameter
Range
None
Return
Float, unit: Hz
Default
None
Menu
Meas > Occupied BW
Example
:MEASure:OBW?
3.7.3.5 Query OBW (:MEASure:OBWidth:OBWidth?)
Command
Format
:MEASure:OBWidth:OBWidth?
Instruction
Use this command to query the occupied bandwidth according to the method
you set.
Query Centroid Result.
Parameter
Type
None
Parameter
Range
None
Return
Float, unit: Hz
Default
None
Menu
Meas > Occupied BW
Example
:MEASure:OBW:OBW?
3.7.3.6 Query OBW Centroid (:MEASure:OBWidth:CENTroid?)
Command
Format
:MEASure:OBWidth:CENTroid?
Instruction
Use this command to query the occupied bandwidth according to the method
you set.
Parameter
Type
None
Parameter
Range
None
Return
Float, unit: Hz