Programming Guide

Table Of Contents
SIGLENT
58 SVA1000X Programming Guide
Instruction
This command returns the value of the channel power density in dBm/Hz.
Parameter
Type
None
Parameter
Range
None
Return
Float
Default
None
Menu
Meas > Ch Power
Example
:MEASure:CHPower:DENSity?
3.7.3 OBW Subsection
3.7.3.1 Select the Method of OBW ([:SENSe]:OBWidth:METHod)
Command
Format
[:SENSe]:OBWidth:METHod PERCent|DBC
[:SENSe]:OBWidth:METHod?
Instruction
This command toggles the method of OBW measurement between percent and
dBc.
Gets the method of OBW measurement.
Parameter
Type
Enumeration
Parameter
Range
PERCent|DBC
Return
Enumeration
Default
PERCent
Menu
Meas > Occupied BW > Meas Setup > Method
Example
:OBW:METHod PERCent
3.7.3.2 Set Percentage(%) Method of OBW
([:SENSe]:OBWidth:PERCent)
Command
Format
[:SENSe]:OBWidth:PERCent <para>
[:SENSe]:OBWidth:PERCent?
Instruction
Edit the percentage of signal power used when determining the occupied
bandwidth. Press {%} to set the percentage ranging from 10.00% to 99.99%.
Gets the percentage of signal power.
Parameter
Type
Float
Parameter
Range
10~99.99
Return
Float
Default
99
Menu
Meas > Occupied BW > Meas Setup > %
Example
:OBW:PERCent 50
3.7.3.3 Set dBc Method of OBW ([:SENSe]:OBWidth:XDB)
Command
Format
[:SENSe]:OBWidth:XDB <value>
[:SENSe]:OBWidth:XDB?
Instruction
Specify the power level used to determine the emission bandwidth as the
number of dB down from the highest signal point, within the occupied bandwidth