Programming Guide

Table Of Contents
SIGLENT
56 SVA1000X Programming Guide
Return
Float, unit: dBm
Default
None
Menu
Meas > ACPR
Example
:MEASure:ACPRatio:LOWer?
3.7.1.7 Query Upper Adjacent Channel Power
(:MEASure:ACPRatio:UPPer:POWer?)
Command
Format
:MEASure:ACPRatio:UPPer:POWer?
Instruction
Return the upper adjacent channel power of ACPR measurement.
Parameter
Type
None
Parameter
Range
None
Return
Float, unit: dBm
Default
None
Menu
Meas > ACPR
Example
:MEASure:ACPRatio:UPPer:POWer?
3.7.1.8 Query Upper Adjacent Channel Power Ratio
(:MEASure:ACPRatio:UPPer?)
Command
Format
:MEASure:ACPRatio:UPPer?
Instruction
Return the upper adjacent channel power to main channel power ratio.
Parameter
Type
None
Parameter
Range
None
Return
Float
Default
None
Menu
Meas > ACPR
Example
:MEASure:ACPRatio:UPPer?
3.7.2 CHP Subsection
3.7.2.1 Integration BW ([:SENSe]:CHPower:BWIDth:INTegration)
Command
Format
[:SENSe]:CHPower:BWIDth:INTegration <freq>
[:SENSe]:CHPower:BWIDth:INTegration?
Instruction
Specifies the integration bandwidth to calculate the power.
Gets the integration bandwidth.
Parameter
Type
Float, unit: Hz, KHz, MHz, GHz
Parameter
Range
100 Hz~1.5 GHz
Zero Span: 0~1.5 GHz
Return
Float, unit: Hz
Default
2 MHz
Menu
Meas > Ch Power > Meas Setup > Integration BW
Example
:CHPower:BWIDth:INTegration 1.0 GHz