Programming Guide

Table Of Contents
SIGLENT
52 SVA1000X Programming Guide
Return
X-axis: Float
Amplitude: Float
Default
X-axis: -1Hz
Amplitude: 0 dBm
Menu
Limit > Limit1|2 Edit
Example
:CALC:LLINe1:DATA 10000000,-20,20000000,-30
3.6.2.9 Add Limit Point Data (:CALCulate:LLINe[1]|2:DATA)
Command
Format
:CALCulate:LLINe[1]|2:ADD <x-axis>,<ampl>
Instruction
Add limit point data
Parameter
Type
X-axis: Float
Amplitude: Float
Parameter
Range
X-axis: 0~1.5GHz
Amplitude: None
Return
X-axis: Float
Amplitude: Float
Default
X-axis: -1Hz
Amplitude: 0 dBm
Menu
Limit > Limit1|2 Edit
Example
:CALCulate:LLINe1:ADD 10000000,-20
3.6.2.10 Delete Assigned Limit Point
(:CALCulate:LLINe[1]|2:DELete)
Command
Format
:CALCulate:LLINe[1]|2:DELete <number>
Instruction
Use this command to delete the assigned limit point.
Parameter
Type
Integer
Parameter
Range
None
Return
None
Default
None
Menu
Limit > Limit1|2 Edit > Del Point
Example
:CALCulate:LLINe1:DELete 2
3.6.2.11 Delete All Limit Points (:CALCulate:LLINe:ALL:DELete)
Command
Format
:CALCulate:LLINe[1]|2:ALL:DELete
Instruction
Use this command to define all the limits points.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Limit > Limit1|2 Edit > Del All
Example
:CALCulate:LLINe2:ALL:DELete