SVA1000X Programming Guide 51
Instruction
Mode sets a limit line to be either an upper or lower type limit line. An upper line
will be used as the maximum allowable value when comparing with the data.
Gets limit type.
Parameter
Type
Enumeration
Parameter
Range
UPPer|LOWer
Return
Enumeration
Default
The default setting of LINe1 is UPPer, the default setting of LINe2 is LOWer
Menu
Limit > Limit1|2 Edit > Type
Example
:CALCulate:LLINe1: TYPE LOWer
3.6.2.6 Limit Mode (:CALCulate:LLINe[1]|2:MODE)
Command
Format
:CALCulate:LLINe[1]|2:MODE LINE|POINt
:CALCulate:LLINe[1]|2:MODE?
Instruction
Sets limit mode
Gets limit mode
Parameter
Type
Enumeration
Parameter
Range
LINE|POINt
Return
Enumeration
Default
LINE
Menu
Limit > Limit1|2 Edit > Mode
Example
:CALCulate:LLINe1: MODE POINt
3.6.2.7 Limit Line Y-axis Value (:CALCulate:LLINe[1]|2:Y)
Command
Format
:CALCulate:LLINe[1]|2:Y <value>
:CALCulate:LLINe[1]|2:Y?
Instruction
Sets the Y-axis value of a limit line. Limit line Y-axis value is set independently
and is not affected by the X-axis units.
Gets the Y-axis value of a limit line.
Parameter
Type
Float
Parameter
Range
-400 dBm~330 dBm
Return
Float
Default
0dBm
Menu
Limit > Limit1|2 Edit > Amplitude
Example
:CALCulate:LLINe1:Y 5dBm
3.6.2.8 Define Limit Points Data (:CALCulate:LLINe[1]|2:DATA)
Command
Format
:CALCulate:LLINe[1]|2:DATA <x-axis>,<ampl>{,<x-axis>, <ampl>}
:CALCulate:LLINe[1]|2:DATA?
Instruction
Use this command to define the limit points.
Gets the defined limit points.
Parameter
Type
X-axis: Float
Amplitude: Float
Parameter
Range
X-axis: 0~1.5GHz
Amplitude: -400 dBm~330 dBm