Programming Guide

Table Of Contents
SIGLENT
44 SVA1000X Programming Guide
Parameter
Range
None
Return
None
Default
None
Menu
Marker > M→Span
Example
:CALCulate:MARKer2:DELTa:SPAN
3.6.1.15 Marker Delta to Center Frequency
(:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:SET]:CENTer)
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:SET]:CENTer
Instruction
This command sets the center frequency equal to the specified delta marker
frequency.
This command can be only used in DELTa|BAND marker mode, Reference
Command:CALCulate:MARKer[1]|2|3|4|5|6|7|8:MODE
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Marker > M→CF
Example
:CALCulate:MARKer3:DELTa:CENTer
3.6.1.16 Peak Search Type
(:CALCulate:MARKer:PEAK:SEARch:MODE)
Command
Format
:CALCulate:MARKer:PEAK:SEARch:MODE MAXimum|MINimum
:CALCulate:MARKer:PEAK:SEARch:MODE?
Instruction
This is for the analyzers internal peak identification routine to recognize a
signal as a peak.
Parameter
Type
Enumeration
Parameter
Range
MAXimum|MINimum
Return
Enumeration
Default
MAXimum
Menu
Peak > Search Config > Peak Type
Example
:CALCulate:MARKer:PEAK:SEARch:MODE MINimum
3.6.1.17 Peak Threshold (:CALCulate:MARKer:PEAK:THReshold)
Command
Format
:CALCulate:MARKer:PEAK:THReshold <value>
:CALCulate:MARKer:PEAK:THReshold?
Instruction
Specifies the minimum signal level for the analyzers internal peak identification
routine to recognize a signal as a peak. This applies to all traces and all
windows.
Gets the minimum signal level for the analyzers internal peak identification
routine to recognize a signal as a peak.
Parameter
Float, unit: dBm