Format
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Instruction
This command sets the per-division display scaling for the y-axis when scale
type of Y axis is set to Log.
Gets Scale/Div when scale type of Y axis is set to Log.
Parameter
Type
Float
Parameter
Range
1 dB ~ 10 dB
Return
Float, unit: dB
Default
10 dB
Menu
Amplitude > Scale/Div
Example
:DISPlay:WINDow:TRACe:Y:PDIVision 10 dB
3.5.3.9 Correction Off ([:SENSe]:CORRection:OFF)
Command
Format
[:SENSe]:CORRection:OFF
Instruction
Turn off the amplitude correction function off and all of the correction sets are
off.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
None
Example
:SENSe:CORRection:OFF
3.5.3.10 Correction Apply State
([:SENSe]:CORRection:CSET:ALL[:STATe])
Command
Format
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Instruction
Turns on or off the amplitude corrections.
When turned on, only the correction sets that were turned on are enabled.
When turned off, all of the correction Sets are disabled. If there is no correction
enabled, state cannot be set to on.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
OFF
Menu
Amplitude > Corrections > Apply Corrections
Example
:SENSe:CORRection:CSET:ALL:STATe OFF
3.5.3.11 Sets Correction X State Off
([:SENSe]:CORRection:CSET[1]|2|3|4:OFF)
Command
Format
[:SENSe]:CORRection:CSET[1]|2|3|4:OFF
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Instruction
Turns the amplitude correction function on/off.