Programming Guide

Table Of Contents
SIGLENT
SVA1000X Programming Guide 29
3.5.3.5 Amplitude OffSets
(:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet )
Command
Format
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet <value>
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet?
Instruction
Sets reference offsets.
Gets reference offsets.
Parameter
Type
Float
Parameter
Range
-300dB~300dB
Return
Float, unit: dB
Default
0dB
Menu
Amplitude > Ref OffSets
Example
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet 2
3.5.3.6 Amplitude Units (:UNIT:POWer)
Command
Format
:UNIT:POWer DBM|DBMV|DBUV|V|W
:UNIT:POWer?
Instruction
Specifies amplitude units for the input, output and display.
Gets amplitude units.
Parameter
Type
Enumeration
Parameter
Range
DBM|DBMV|DBUV|DBUA|V|W,
Return
Enumeration
Default
DBM
Menu
Amplitude > Units
Example
:UNIT:POWer DBMV
3.5.3.7 Scale Type
(:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing)
Command
Format
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing LINear|LOGarithmic
:DISPlay:WINDow:TRACe:Y[:SCALe]:SPACing?
Instruction
Toggles the vertical graticule divisions between logarithmic unit and linear unit.
The default logarithmic unit is dBm, and the linear unit is V.
Gets scale type.
Parameter
Type
Enumeration
Parameter
Range
LINear|LOGarithmic
Return
Enumeration
Default
LOGarithmic
Menu
Amplitude > Scale Type
Example
:DISPlay:WINDow:TRACe:Y:SPACing LINear
3.5.3.8 Scale/Div (:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision)
Command