Programming Guide

Table Of Contents
SIGLENT
28 SVA1000X Programming Guide
Return
Float, unit: dBm
Default
0 dBm
Menu
Amplitude > Ref Level
Example
:DISPlay:WINDow:TRACe:Y:RLEVel 20 DBM
3.5.3.2 Input Attenuator ([:SENSe]:POWer[:RF]:ATTenuation)
Command
Format
[:SENSe]:POWer[:RF]:ATTenuation <value>
[:SENSe]:POWer[:RF]:ATTenuation?
Instruction
Sets the input attenuator of the spectrum analyzer.
Gets the input attenuator.
Parameter
Type
Integer
Parameter
Range
0 dB ~ 31 dB
Return
Integer, unit: dB
Default
20 dB
Menu
Amplitude > Attenuator
Example
:POWer:ATTenuation 10
3.5.3.3 Attenuator Auto Mode
([:SENSe]:POWer[:RF]:ATTenuation:AUTO)
Command
Format
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Instruction
This command turns on/off auto input port attenuator state.
Gets input port attenuator state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
ON
Menu
Amplitude > Attenuator
Example
:POWer:ATTenuation:AUTO?
3.5.3.4 Preamp on-off ([:SENSe]:POWer[:RF]:GAIN[:STATe])
Command
Format
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Instruction
Turns the internal preamp on/off.
Gets preamp on-off state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
OFF
Menu
Amplitude > Preamp
Example
:POWer:GAIN ON