Programming Guide

Table Of Contents
SIGLENT
SVA1000X Programming Guide 27
Default
None
Menu
Span> Zoom In
Example
:FREQuency:SPAN:HALF
3.5.1.12 Zoom Out ([:SENSe]:FREQuency:SPAN:DOUBle)
Command
Format
[:SENSe]:FREQuency:SPAN:DOUBle
Instruction
Sets the frequency span to double the previous span setting.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Span> Zoom Out
Example
:FREQuency:SPAN:DOUBle
3.5.2 Auto Tune Subsection
3.5.2.1 Auto Tune ([:SENSe]:FREQuency:TUNE:IMMediate)
Command
Format
[:SENSe]:FREQuency:TUNE:IMMediate
Instruction
Auto tune the spectrum analyzer parameter to display the main signal.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Auto Tune
Example
:FREQuency:TUNE:IMMediate
3.5.3 Amplitude Subsection
3.5.3.1 Reference Level
(:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel)
Command
Format
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <value>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Instruction
This command sets the reference level for the Y-axis.
Gets reference level.
Parameter
Type
Float, unit: dBm, dBmV, dBuV, V, W
Parameter
Range
Unit is dBm: -100 dBm ~ 30 dBm
Unit is dBmV: -53.01 dBmV ~ 76.99 dBmV,
Unit is dBuV: 6.99 dBuV ~ 136.99 dBuV,
Unit is Volts: 2.24 uV ~ 7.07 V
Unit is Watts: 100 fW ~ 1 W