Data Sheet

SVA1000X DataSheet 9
Tracking Generator
Frequency Parameter
Frequency Range
SVA1015X
SVA1032X
5 MHz~1.5 GHz
100 kHz~3.2 GHz
Frequency resolution
1 Hz, Zero Span
RBW
100 Hz ~ 1 MHz, sweep mode
Power Parameter
Output level
-20 dBm ~ 0 dBm
Output level resolution
1 dB
Output flatness
+/-3 dB (nom.)
Normalization Trace
Ref A/B/C->D
VSWR
< 2 (nom.)
Connector and Impendence
N-type female, 50 Ω
Average safe reverse power
Total : 30 dBm (1 W)
Maximum safe reverse level
Voltage: ± 50 V
DC
EMI Filter and Quasi-Peak Detector Kit (Option SVA1000X-DTF)
Measurement
EMI filter RBW(-6dB)
200 Hz, 9 kHz, 120 kHz, 1MHz (following CISPR 16-1-1)
Detector
Peak, Average, RMS, Quasi-peak(following CISPR 16-1-1)
QPD Dwell time
0 us ~ 10 s
PC Application Software
EasySpectrum EMI pre-compliance test Software
Frequency axis
Linear, Logarithmic
Advanced Measurement Kit (Option SVA1000X-AMK)
Power Measurement
CHP, Channel Power
Channel Power, Power Spectral Density
ACPR, Adjacent Channel
Power Ratio
Main CH Power, Left channel power, Right channel power
OBW, Occupied Bandwidth
Occupied Bandwidth, Transmit Frequency Error
T-Power, Time Domain Power
Zero Span Integrated Power
CNR, Carrier Noise Ratio
C/N, Noise Power
Non-Linear Measurement
Harmonic measurement
Max Harmonic number 10
TOI, Third-Order Intercept
Measure the third-order products and intercepts from two tones
Spectrum Monitor Measurement
Spectrogram