SVA1000X Series Spectrum Analyzers
SVA1000X Spectrum Analyzer Data Sheet Features and Benefits SVA1015X All-Digital IF Technology Frequency Range from 9 kHz up to 1.5 GHz -156 dBm/Hz Displayed Average Noise Level (Typ.) -99 dBc/Hz @10 kHz Offset Phase Noise (1 GHz, Typ.) Level Measurement Uncertainty < 1.2 dB (Typ.) 1 Hz Minimum Resolution Bandwidth (RBW) General Description The SIGLENT SVA1000X series spectrum analyzers Preamplifier Standard 1.
SVA1000X Spectrum Analyzer Data Sheet Model and Main index Model SVA1015X Frequency Range 9 kHz~1.5 GHz Resolution Bandwidth 1 Hz~1 MHz Displayed Average Noise Level -156 dBm/Hz Phase Noise <-99 dBc/Hz@1 GHz, 10 kHz offset Total Amplitude Precision ≤1.2 dB Touch Screen Standard Tracking Generator Standard Vector Network Analysis S11, S21 Distance To Fault 10 MHz-1.
SVA1000X Spectrum Analyzer Data Sheet Vector Network Analysis Mode Distance to Fault Mode ASK/FSK Digital Modulation Analysis Mode EMI filter and Quasi-peak Detector in EMI Precompliance Test Kit ACPR in Advanced Measrement Kit Spectrum Monitor in Advanced Measrement Kit Easy Spectrum Software in EMI Pre-compliance Test Kit Remote Control through Web Browser WWW.SIGLENT.
SVA1000X Spectrum Analyzer Data Sheet Utility Kit Near Field Probe Set SRF5030 USB-GPIB Adaptor Soft Carrying Bag Near Field Probe Set SRF5030T Mechanical Calibration Kit Specifications Specifications are valid under the following conditions: The instrument is within the calibration period, has been stored between 0 and 50°C for at least 2 hours prior to use, and has been powered on and warmed up for at least 40 minutes. The specifications include the measurement uncertainty, unless otherwise noted.
SVA1000X Spectrum Analyzer Data Sheet Amplitude Characteristic Amplitude and Level Measurement range DANL to +10 dBm, 100 kHz~1 MHz, preamplifier off DANL to +20 dBm, 1 MHz~1.5 GHz, preamplifier off Reference level -100 dBm to +30 dBm, 1 dB steps Preamplifier 20 dB (nom.), 9 kHz~1.
SVA1000X Spectrum Analyzer Data Sheet Amplitude Characteristic Distortion and Spurious Responses Second harmonic distortion -65 dBc (nom.) fc≥50 MHz, mixer level -30dBm, attenuation = 0dB, preamp off, 20 °C to 30 °C Third-order intercept +8 dBm (typ.) fc≥50 MHz, two -20 dBm tones at input mixer spaced by 100 kHz, attenuation = 0 dB, preamp off, 20 °C to 30 °C 1dB Gain Compression >-5 dBm (nom.
SVA1000X Spectrum Analyzer Data Sheet Digital Modulation Analysis Mode Frequency range 5 MHz to 1.5 GHz Carrier Power Accuracy ±2 dB, nominal Carrier Power Range -30 dBm to +20 dBm, nominal ASK Symbol rate range 1 kHz to 100 kHz Modulation depth/index range 5% to 95% Accuracy ±4%, nom. FSK Symbol rate range (β = deviation/Symbol rate) 100 Hz to 20 kHz 1≤β≤20 25 kHz to 50 kHz 1≤β≤8 50 kHz to 100 kHz 1≤β≤4 FSK deviation 1 kHz to 400 kHz Accuracy ±4%, nom.
SVA1000X Spectrum Analyzer Data Sheet Ordering Information Product Description SVA1000X Product Code Spectrum Analyzer, 9 kHz~1.
SVA1000X Series Spectrum Analyzer About SIGLENT SIGLENT is an international high-tech company, concentrating on R&D, sales,production and services of electronic test & measurement instruments. SIGLENT first began developing digital oscilloscopes independently in 2002.