User Manual
SIGLENT
SSG3000X User Manual 80
Chapter 3 Application Examples
3.1 Output RF signal
Output a RF signal with a frequency of 3GHz and amplitude of 0dBm from the RF
OUTPUT 50 connector.
1. Restore factory settings
A) Press UTILITY or the touch screen to set System -> Setting -> Preset Type->
Default.
B) Press PRESET, or reset the system settings to perform reset operations. Make the
frequency offset and the phase defaults 0.
2. Frequency setting
Press FREQ, and enter the 3GHz with the numeric keypad, or click the frequency on the
touch screen to pop up the keypad to input 3GHz.
3. Level setting
Press LEVEL, and enter the 0dBm with the numeric keypad, or click the amplitude on the
touch screen to pop up the keypad to input 0dBm on the keypad.
4. RF output
Press RF ON/OFF or press the HOME key back to the main interface, select the RF block,
the white square on the left of ON to enable RF output. When enabled, the RF ON/OFF
key backlight turns green and the RF state in the status bar changes to blue.