User Manual
SIGLENT
SSG3000X User Manual 63
finally press ENTER to make the setup effective.
2.6.1.4 Self Test
1. Press Utility ->Self test to enter the self test menu, including:LCD Test
SSG3000X cycles through the red, green and blue pixels of the display to verify whether
there are dot defects on the screen.
Press UTILITY->Self-Test ->LCD Test, then press 7 to change the color and press 8 to
exit.
2. Key Test
PressUTILITY->Self-Test ->Key Test, to enter the keyboard test interface. Press the
function keys at the front panel one by one and observe whether the corresponding key is
checked. If not, an error may have occurred in that key. To exit the test, press 8 three
times or click the screen.
3. LED Test
Press UTILITY->Self-Test ->LED Test, to enter the LED test interface. Press 7 to light or
extinguish the key light of the button MODON/OFF and RFON/OFF and press 8 to exit
the test.
4. Board Test
Press UTILITY->Self-Test ->LEDTest, to enter the board test interface. Test whether
writing and reading are normal for CPLD and FPGA.
5. Touch Screen Test
Press UTILITY->Self-Test ->TouchScreenTest, to enter the touch screen test interface.
Click the marks on the interface to test the touch screen and press 8 to exit the test.