User Manual
SIGLENT
SSG3000X User Manual 47
mode is ―Single‖ or ―Double‖.
2.4.4.7 Pulse Width
The pulse width indicates the duration when the pulse signal is at a high level. The width
of the first pulse of the single pulse modulated signal or the double pulse modulated signal
can be set by touching screen keyboard or pressing the front panel keypad.
Note: This setting item will be hidden when the modulation source is ―Ext‖ or the pulse
mode is ―Train‖.
2.4.4.8 Double Pulse Delay
The double pulse delay indicates the delay from the start of the first pulse to the start of
the second pulse in a single cycle of the double pulse modulated signal. This parameter
can be set by touching screen keyboard or pressing the front panel keypad.
Note: This setting item will be hidden when the modulation source is ―Ext‖ or the pulse
mode is ―Train‖.
2.4.4.9 #2 Pulse Width
#2 Pulse Widthrepresents the width of the second pulse of the double pulse modulated
signal. This parameter can be set by touching screen keyboard or pressing the front panel
keypad.
Note: This setting item will be hidden when the modulation source is ―Ext‖ or the pulse
mode is ―Train‖.