2.3 Sweep Setting
When the RF sweep function is enabled, the RF signal is output from the N-type
connector of [RF OUTPUT 50Ω] on the front panel.
Note: The RF signal will be on only when the RF switch is on.
Press SWEEP to enter RF sweep setting.
2.3.1 Sweep State
The sweep state is off by default. The signal generator SSG3000X offers three different
RF sweep types: Freq, Level, and Freq& Level.
Clicking the drop-down box of the sweep state to choose a sweep type:
Off: Default state. The sweep disabledClose RF sweep function.
Freq: Enable frequency sweep. The frequency parameter bar refreshes the current
frequency value in real time and the current sweep progress can be observed on the
frequency sweep progress bar.
Level: Enable level sweep. The frequency parameter bar refreshes the current level
value in real time and the current sweep progress can be observed on the level
sweep progress bar.
Freq& Level: Enable frequency sweep and level sweep function. The frequency and
level parameter bar refreshes the current frequency and level value in real time and
the current sweep progress can be observed on the frequency and level sweep
progress bar.
2.3.2 Step Sweep
The step sweep state is on by default. Click the slide to switch the on-off state. Click the
to enter the parameter setting menu of step sweep.
Note: The step sweep and list sweep are mutually exclusive. One will shut off