SSG3000X User Manual 21
2. Touch screen operation
Click SHORTCUT->FREQ-> Freq Offset to enter the interface of frequency offset setting.
Click the frequency offset parameter, and then pop up the keypad interface, then click
the number symbol in the keypad to change the value. Click the unit to modify the unit.
Click the check mark to confirm the modified value and selected unit.
Long press the edit box to enter the edit mode. Press the up and down direction keys,
rotate the knob or press the numeric keypad to modify the value, and press the
ENTER key, knob or ESC key to exit the edit mode.
Note:
The relationship between the RF output frequency, the display frequency and
frequency offset is:
Display frequency = RF output Frequency + Frequency offset
The difference value between the display frequency and the frequency offset is the
actual RF output frequency, which must not exceed the instrument frequency output
range.
2.1.3 Phase Offset
Setting the phase offset from the RF signal
Application:When multiple RF sources are outputting RF signals at the same time,
multiple signals can be output to the same phase or a fixed phase offset by adjusting this
parameter.
Also you should set the devices the same frequency and synchronous them.
1. Keypad operation
Press FREQ-> Phase Offset to make the focus fall on this parameter item.
Use the numeric keypad to input the desired value and press the unit button to select
the desired unit. Press ENTER to select the current unit.