3.3 Output a pulse train
A pulse train is a user-defined sequence of high and low RF levels with specific time
durations for each part. In this example, we will show how to configure the RF source to
output the following pulse train:
Serial
number
Positive pulse
width
Negative pulse
width
Repetitions
1
10ms
30ms
1
2
20ms
20ms
2
1. Restore the factory default settings
A) Press UTILITY or the touch Preset Type-> Default.
B) Press PRESET or reset the system settings to perform the reset operation. Make the
frequency offset and the phase defaults 0.
2. Install thepulse train option
Activate the pulse train generator option.
3. Edit pulse list
A) Press MOD or select MOD from the Home page. You can swipe left or right or press
MOD repeatedly to reach the Pulse Mode screen. Set Pulse Mode to Train, and the Pulse
Train menu will appear.
B) Click in Pulse Train menu to enter settings, set the positive width of first pulse to
10ms, the negative width to 30ms, and the count to 1.
C) Click the + on the left to add one line, set the positive pulse width to 20ms, the