2.7.1.6 Auto Zero
When the power meter is zeroed, the zero measurement deviation and noise influence
can be reduced, and the accuracy of the RF power measurement can be improved.
Disable: The zero buttons will be hidden.
Internal Zeroing: In this way, the power sensor remains connected to the RF signal
during the zeroing process. Zero buttons is displayed.
External Zeroing: When this mode is selected, you must remove the RF signal to be
measured before the power sensor can be zeroed. Otherwise it will affect the zeroing
result. Zero buttons is displayed.
Press the ―Click to perform zeroing‖ button, and the instrument performs the zero
calibration operation of the power meter according to the ―zero calibration type‖ you
selected. This feature can be used to reduce the effects of noise and zero offset on the
measurement results. And the button name will change from ―Click to perform zeroing‖ to
―Zero In Progress...‖ After the zero adjustment is completed, the button name will be
restored to the word ―Click to perform zeroing‖.
Note: If the power meter does not have an internal or external zeroing option, the ―internal‖
and ―external‖ in the drop-down boxes are replaced by the ―Enable‖ option.
2.7.1.7 Measure Frequency
The default is ―Auto‖. Click the drop down box to switch the mode.
Auto: In this mode, the instrument configures the frequency value based on the
output frequency of the RF.
Manual: In this mode, the measurement frequency can be set by the touch screen
keyboard or the front panel keypad.