User Manual

SIGLENT
SSG3000X User Manual 57
instrument waits for the next trigger event.
3. Ext
Generate a single sweep cycle when an external trigger event occurs.
2.5.2.9 Sweep Shape
Select LF->LF Sweep->Sweep Shape to select the waveform shape of the sweep signal.
1. Sawtooth
The sweep runs from the start to the stop frequency. Each subsequentsweep starts at the
start frequency that means the shape ofthe sweep sequence resembling a Sawtooth.
2. Triangle
The sweep runs from start to stop frequency and back, that meansthe shape of the sweep
resembles a triangle. Subsequent sweepstarts at the initial start frequency.
2.5.2.10 Sweep Space
Select LF->LF Sweep->Sweep Space to select the mode for the calculation of the
frequency sweep intervals.
1. Linear
Take the frequency value entered as an absolute value in Hz
2. Logarithmic
Takes the value entered as a logarithmic value that means as a constantfraction of
current frequency in %.