Press ENTER or press the knob to enter the parameter editing state, and then use
the left and right direction keys to move the cursor to the desired position. Press the
up and down direction keys, rotate the knob or press the numeric keypad to modify
the value. Press ENTER, knob or ESC to exit the editing mode.
2. Touch operation
Click on SHORTCUT->LEVEL->Level Offset to enter the interface of level offset setting.
Click the level parameter, then pop up the keypad interface, then click the number
symbol in the keypad to change the value. Click the check mark to confirm the
modified value.
Long press edit box to enter the edit mode. Press the up and down direction keys,
rotate the knob or press the numeric keypad to modify the value, and press the
ENTER key, knob or ESC key to exit the edit mode.
Note:
The difference value between the display level and the level offset is the actual RF
output level, which must not exceed the instrument level output range.
2.2.3 ALC State
Setting the ALC’s working state
ALC is Automatic Level Control .It is an adaptive control system to stabilize the RF output
level.
It continuously monitors the current level and adjusts it to keep a steady state over
temperature and time.