User Manual

SIGLENT
SSG3000X User Manual 104
If the current scan type is set to Frequency, the amplitude value does not change.
5 USB device cannot be recognized:
(1) Check if the USB device is connected to other instruments or the computer can work
normally.
(2) Confirm the flash type U disk device used. This instrument does not support the hard
disk type U disk device.
(3) After restarting the instrument, insert the USB flash drive to check.
(4) If you are unable to use the USB flash drive normally, please contact.
6 Inaccurate measurement results or insufficient accuracy:
Users can obtain detailed descriptions of technical indicators from the back of this manual
to calculate system errors and check measurement results and accuracy issues. To
achieve the performance metrics listed in this manual, you need:
(1) Check if the RF signal source is within the calibration period (1 year).
(2) Verify that the RF signal source has been warmed up for at least 30 minutes before
testing.
(3) Check that the performance of the test equipment used meets the requirements.
(4) Ensure that the test equipment used is within the calibration cycle.
(5) Check if the test equipment used is under the working conditions required by its
manual.
(6) Check if all connections are tightened.
7 Pop-up Message:
The instrument may display prompt messages, error messages or state messages
according to the current working status. These messages are displayed to help you to use
the instrument correctly and are not instrument failures.