Data Sheet
SSG3000X RF Generator datasheet
WWW.SIGLENT.COM
SSG3000X RF Generator datasheet
SPECIFICATIONS
Specications are valid under the following condition: The instrument is within the calibration period, has been stored between 0 and 50
℃
for at least 2 hours prior to
use, and has been powered on and warmed up for at least 40 minutes. The specications include the measurement uncertainty, unless otherwise noted.
Specications: All products are guaranteed to meet published specications when operating temperatures from 5 to 45
℃
unless otherwise noted.
Typical(typ.): Performance deemed typical implies that 80 percent of the measurement results will meet the typical published performance with a 95th percentile
condence level at room temperature (approximately 25
℃
). Typical performance is not warranted and does not include measurement uncertainty.
Nominal(nom.): This value indicate the expected mean or average performance, or an attribute whose performance is by design, such as the 50 ohm connector.
Frequency characteristics
Frequency
Frequency range
SSG3032X CW MODE 9 kHz~3.2 GHz
SSG3021X CW MODE 9 kHz~2.1 GHz
SSG3032X-IQE CW MODE 9 kHz~3.2 GHz IQ MODE 10 MHz~3.2 GHz
SSG3021X-IQE CW MODE 9 kHz~2.1 GHz IQ MODE 10 MHz~2.1 GHz
Frequency resolution 0.01 Hz
Setting time
<5 ms (typ.), ALC ON
<10 ms (typ.), ALC OFF (S&H)
Resolution of phase offset
setting
0.1°
Frequency Band
[1]
Band Frequency range N
1 9 kHz≤f≤1 MHz 0.25
2 1 MHz<f≤250 MHz 0.5
3 250 MHz<f≤500 MHz 0.125
4 500 MHz<f<1000 MHz 0.25
5 1000 MHz≤f<2000 MHz 0.5
6 2000 MHz≤f≤3200 MHz 1
[1] N is a factor used to help dene certain specications with the document
Frequency Reference
Reference frequency 10.000000 MHz
Initial calibration accuracy <0.2 ppm
Temperature stability <1 ppm/year, 0
℃
~50
℃
Frequency aging rate <0.5 ppm/rst year, 3.0 ppm/20 years
Frequency sweep
Sweep type
frequency step (linear or logarithmic step)
arbitrary list
Sweep range full frequency range
Sweep sheep triangle, saw-tooth
Sweep mode single, continuous
Step spacing linear, logarithmic
Number of points
step sweep 2~65535
list sweep 2~500
Dwell time range 10 ms~100 s
Dwell time setting resolution 0.1 ms
Trigger source auto, keyboard, external connector, bus
Trig slop positive, negative when trigger source is external
,