Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
62 SSA3000X Plus Programming Guide
Parameter
Range
None
Return
None
Default
None
Menu
Marker Fn > Freq Counter
Example
:CALCulate:MARK:FCOunt:X?
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:RESult?
Instruction
Gets the result of N dB bandwidth measurement.
Parameter
Type
None
Parameter
Range
None
Return
Float
Default
None
Menu
Marker Fn > N dB BW
Example
:CALCulate:MARK1:BANDwidth:RESult?
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:NDB <value>
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:BANDwidth:NDB?
Instruction
Sets the reference value of N dB bandwidth measurement.
Gets the reference value of N dB bandwidth measurement.
Parameter
Type
Float
Parameter
Range
-100dB ~ 100dB
Return
Float
Default
-3 dB
Menu
Marker Fn > N dB BW
Example
:CALCulate:MARK1:BANDwidth:NDB 10
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:READout
FREQuency|TIME|PERiod
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:X:READout?
Instruction
Toggles the marker X-Axis readout between frequency, time and period.
Gets the marker X-Axis readout type.
Parameter
Type
Enumeration
Parameter
Range
FREQuency|TIME|PERiod
Return
Enumeration