Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
60 SSA3000X Plus Programming Guide
Default
None
Menu
Peak > Next Peak
Example
:CALCulate:MARKer1:MAXimum:NEXT
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:LEFT
Instruction
Places the selected marker on the next highest signal peak to the left of the
current marked peak.
(based on the search mode settings, include: peak search mode, peak
threshold and peak excursion, Reference Commands:
:CALCulate:MARKer:PEAK:SEARch:MODE
:CALCulate:MARKer:PEAK:THReshold
:CALCulate:MARKer:PEAK: EXCursion)
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Peak > Next Left Peak
Example
:CALCulate:MARKer1:MAXimum:LEFT
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:MAXimum:RIGHt
Instruction
Places the selected marker on the next highest signal peak to the right of the
current marked peak.
(based on the search mode settings, include: peak search mode, peak
threshold and peak excursion, Reference Commands:
:CALCulate:MARKer:PEAK:SEARch:MODE
:CALCulate:MARKer:PEAK:THReshold
:CALCulate:MARKer:PEAK: EXCursion)
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Peak > Next Right Peak
Example
:CALCulate:MARKer1:MAXimum:RIGHt
Command
Format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch
Instruction
Positions a pair of delta markers on the highest and lowest points on the
trace.