Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
58 SSA3000X Plus Programming Guide
Default
-160.0 dBm
Menu
Peak > Search Config > Peak Threshold
Example
:CALCulate:MARKer:PEAK:THReshold -50
Command
Format
:CALCulate:MARKer:PEAK:EXCursion <value>
:CALCulate:MARKer:PEAK:EXCursion?
Instruction
Specifies the minimum signal excursion above the threshold for the internal
peak identification routine to recognize a signal as a peak.
Parameter
Type
Float, unit: dB
Parameter
Range
0 ~ 200.0dB
Return
Float, unit: dB
Default
0 dB
Menu
Peak > Search Config > Peak Excursion
Example
:CALCulate:MARKer:PEAK:EXCursion 10
Command
Format
:CALCulate:MARKer:PEAK:TABLe ON|OFF|0|1
:CALCulate:MARKer:PEAK:TABLe?
Instruction
Toggles the peak table between on and off.
Gets the status of the peak table.
Parameter
Type
Boolean
Parameter
Range
ON|OFF|0|1
Return
0|1
Default
0
Menu
Peak > Peak Table
Command
Format
:CALCulate:PEAK:TABLe?
Instruction
Get peak table data.
Parameter
Type
None
Parameter
Range
None
Return
String
Default
None
Menu
Peak > Peak Table
Example
:CALCulate:PEAK:TABLe?