Programming Guide

Table Of Contents
SIGLENT
46 SSA3000X Plus Programming Guide
:CALCulate[:SELected]:MATH:FUNCtion
:TRACe:MATH:X
:TRACe:MATH:Y
:TRACe:MATH:Z
:TRACe:MATH:OFFSet
[:SENSe]:FREQuency:TUNE:IMMediate
[:SENSe]:DETector:TRACe[1]|2|3|4[:FUNCtion]
[:SENSe]:AVERage:TYPE
[:SENSe]:AVERage:TRACe[1]|2|3|4:COUNt
[:SENSe]:AVERage:TRACe[1]|2|3|4:CLEar
Command
Format
:TRACe[1]|2|3|4:MODE
WRITe|MAXHold|MINHold|VIEW|BLANk|AVERage
:TRACe[1]|2|3|4:MODE?
Instruction
Selects the display mode for the selected trace.
Parameter
Type
Enumeration
Parameter
Range
WRITe: puts the trace in the normal mode, updating the data.
MAXHold: displays the highest measured trace value for all the data that has
been measured since the function was turned on.
MINHold: displays the lowest measured trace value for all the data that has
been measured since the function was turned on.
VIEW: turns on the trace data so that it can be viewed on the display.
BLANk: turns off the trace data so that it is not viewed on the display.
AVERage: averages the trace for test period.
Return
Enumeration
Default
Trace1: WRITe, Trace2|3|4: BLANk
Menu
Trace
Example
:TRAC1:MODE VIEW
Command
Format
:TRACe[:DATA]? 1|2|3|4
Instruction
This query command returns the current displayed data.
You can also add trace parameters directly after trace as
:TRACe[1]|2|3|4[:DATA]?
Parameter
Type
Enumeration
Parameter
Range
1|2|3|4 or A|B|C|D or TRACE1| TRACE2| TRACE3| TRACE4
Return
String
Default
1
Menu
None