Programming Guide

Table Of Contents
SIGLENT
SSA3000X Plus Programming Guide 41
Default
None
Menu
Example
:INITiate:IMMediate
Command
Format
:INITiate:RESTart
Instruction
Restart the current sweep.
:INITiate:RESTart and :INITiate:IMMediate perform exactly the same
function.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Example
:INITiate:RESTart
Command
Format
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Instruction
Sets continuous sweep mode on-off.
Gets continuous sweep mode state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
ON
Menu
Sweep > Sweep
Example
:INITiate:CONTinuous OFF
ABORt
This command is used to stop the current measurement. It aborts the
currentmeasurement as quickly as possible, resets the sweep and trigger
systems, and puts the measurement into an "idle" state.
If the analyzer is set for Continuous measurement, it sets up the
measurement and initiates a new data measurement sequence with a new
data acquisition (sweep)taken once the trigger condition is met.
If the analyzer is set for Single measurement, it remains in the "idle" state
until an :INIT:IMM command is received.