SSA3000X Plus Programming Guide 39
Instruction
Sets sweep mode.
Gets sweep mode.
Parameter
Type
Enumeration
Parameter
Range
AUTO|FFT|SWEep
Return
Enumeration
Default
SWEep
Menu
Sweep
Example
:SWEep:MODE SWEep
Command
Format
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Instruction
Specifies the time in which the instrument sweeps the display. A span value
of 0 Hz causes the analyzer to enter zero span mode. In zero span the X-
axis represents time rather than frequency.
Parameter
Type
Float, unit: ks, s, ms, us
Parameter
Range
450us ~ 1500 s
Return
Float, unit: s
Default
312.416ms(216.288ms, 192.256ms, 168.224ms, 120.160ms)
Menu
Sweep > Sweep Time
Example
:SWEep:TIME 5s
Command
Format
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Instruction
This command turns on/off auto sweep time state.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
ON
Menu
Sweep > Sweep Time
Example
:SWEep:TIME:AUTO ON
Command
Format
[:SENSe]:SWEep:SPEed NORMal|ACCUracy
[:SENSe]:SWEep:SPEed?
Instruction
Toggles the sweep speed between normal and accuracy.
Parameter
Type
Enumeration