Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
SSA3000X Plus Programming Guide 37
Command
Format
[:SENSe]:CORRection:CSET[1]|2|3|4:ADD <x1,y1,x2,y2;…>
Instruction
Add Correction Points
Parameter
Type
String<freq, ampl,freq, ampl,freq, ampl,……>
Parameter
Range
None
Return
Default
None
Menu
Amplitude > Corrections > CorrectionX > Add Point
Example
:CORRection:CSET2:ADD 10000000,-10,15000000,-12
Command
Format
[:SENSe]:CORRection:CSET[1]|2|3|4:DELete <index>
Instruction
Delete Correction Points
Parameter
Type
Serial number of Correction Points
Parameter
Range
None
Return
Default
None
Menu
Amplitude > Corrections > CorrectionX > Del Point
Example
:CORRection:CSET2: DELete 2
Command
Format
[:SENSe]:CORRection:CSET[1]|2|3|4:ALL:DELete
Instruction
Add All Correction Points
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Amplitude > Corrections > CorrectionX > Del All
Example
:CORRection:CSET2: ALL:DELete
Command
Format
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA <x1,y1,x2,y2;…>
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?