Programming Guide

Table Of Contents
SIGLENT
SSA3000X Plus Programming Guide 31
:FREQuency:OFFSet 1 GHz
Command
Format
[:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
Instruction
Sets the frequency span. Setting the span to 0 Hz puts the analyzer into zero
span.
Gets span value.
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
0 Hz, 100 Hz ~ 3.2GHz
Return
Float, unit: Hz
Default
1.5 GHz
Menu
Span > Span
Example
:FREQuency:SPAN 1 GHz
Command
Format
[:SENSe]:FREQuency:SPAN:FULL
Instruction
Sets the frequency span to full scale.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Span > Full Span
Example
:FREQuency:SPAN:FULL
Command
Format
[:SENSe]:FREQuency:SPAN:ZERO
Instruction
Sets the frequency span to zero span.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Span > Zero Span
Example
:FREQuency:SPAN:ZERO