Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
30 SSA3000X Plus Programming Guide
Return
Float, unit: Hz
Default
320 MHz
Menu
Frequency > Freq Step
Example
:FREQuency:CENTer:STEP 2 MHz
Command
Format
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Instruction
Specifies whether the step size is set automatically based on the span.
Gets center frequency step mode.
Parameter
Type
Boolean
Parameter
Range
OFF|ON|0|1
Return
0|1
Default
ON
Menu
Frequency > Freq Step
Example
:FREQuency:CENTer:STEP:AUTO OFF
Command
Format
[:SENSe]:FREQuency:CENTer:SET:STEP
Instruction
Sets step value equal to center frequency.
Parameter
Type
None
Parameter
Range
None
Return
None
Default
None
Menu
Frequency > CF→Step
Example
:FREQuency:CENTer:SET:STEP
Command
Format
[:SENSe]:FREQuency:OFFSet <freq>
[:SENSe]:FREQuency:OFFSet?
Instruction
Sets the frequency offset of the spectrum analyzer.
Gets the frequency offset.
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
-100 GHz ~ 100 GHz
Return
Float, unit: Hz
Default
0 Hz
Menu
Frequency > Freq Offset