Programming Guide

Table Of Contents
SIGLENT
SSA3000X Plus Programming Guide 29
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
50 Hz~3.199999950 GHz
Zero Span: 0~3.2 GHz
Return
Float, unit: Hz
Default
1.6 GHz
Menu
Frequency > Center Freq
Example
:FREQuency:CENTer 0.2 GHz
Command
Format
[:SENSe]:FREQuency:STARt <freq>
[:SENSe]:FREQuency:STARt?
Instruction
Sets the start frequency of the spectrum analyzer.
Gets the start Frequency.
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
50 Hz~3.199999950 GHz
Zero Span: 0~3.2 GHz
Return
Float, unit: Hz
Default
0 Hz
Menu
Frequency > Start Freq
Example
:FREQuency:STARt 100 Hz
Command
Format
[:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Instruction
Sets the stop frequency of the spectrum analyzer.
Gets the stop frequency.
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
50 Hz~3.199999950 GHz
Zero Span: 0~3.2 GHz
Return
Float, unit: Hz
Default
1.5 GHz
Menu
Frequency > Stop Freq
Example
:FREQuency:STOP 1.0 GHz
Command
Format
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Instruction
Specifies the center frequency step size.
Gets the center frequency step.
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz
Parameter
Range
1 Hz~3.2 GHz