Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
SSA3000X Plus Programming Guide 101
Parameter
Range
4, 6, 8, 10, 12, 14, 16
Return
Discrete
Default
4
Menu
Meas > Points/Symbol
Example
DDEMod:SYMBol:POINts 14
Command
Format
[:SENSe]:DDEMod[:FORMat]:RLENgth <integer>
[:SENSe]:DDEMod[:FORMat]:RLENgth?
Instruction
Sets meas length
Gets meas length
Parameter
Type
Integer
Parameter
Range
16 ~ 4096
Return
Integer
Default
128
Menu
Meas > Meas Length
Example
:DDEMod:RLENgth 200
Command
Format
[:SENSe]:DDEMod:FILTer[:MEASurement]
[:SENSe]:DDEMod:FILTer[:MEASurement]?
Instruction
Sets meas filter
Gets meas filter
Parameter
Type
Enumeration
Parameter
Range
OFF
RRCosine
RCOSine
GAUSsian
HSIN
Return
0|1
Default
ASK, FSK, PSK, QAM Default is RCOSine
MSK Default is OFF
Menu
Meas > Filter Setup > Meas Filter
Example
:DDEMod:FILTer HSIN
Command
Format
[:SENSe]:DDEMod:FILTer:REFerence
[:SENSe]:DDEMod:FILTer:REFerence?
Instruction
Sets reference filter
Gets reference filter
Parameter
Type
Enumeration