SSA3000X Series DataSheet-2015.
SSA3000X Spectrum Analyzer Data Sheet Features and Benefits SSA3032X SSA3021X All-Digital IF Technology Frequency Range from 9 kHz up to 3.2 GHz -161 dBm/Hz Displayed Average Noise Level (Typ.) -98 dBc/Hz @10 kHz Offset Phase Noise (1 GHz, Typ.) Total Amplitude Accuracy < 0.7 dB 10 Hz Minimum Resolution Bandwidth (RBW) Standard Preamplifier General Description Siglent’s SSA3000X series of spectrum analyzers Up to 3.2 GHz Tracking Generator Kit (Opt.) Reflection Measurement Kit (Opt.
SSA3000X Spectrum Analyzer Data Sheet Model and Main index Model SSA3032X SSA3021X Frequency Range 9 kHz~3.2 GHz 9 kHz~2.1 GHz Resolution Bandwidth 10 Hz~1 MHz, in 1-3-10 sequence 10 Hz~1 MHz, in 1-3-10 sequence Displayed Average Noise Level -161 dBm/Hz, Normalize to 1 Hz (typ.) -161 dBm/Hz, Normalize to 1 Hz (typ.) Phase Noise <-98 dBc/Hz@1 GHz, 10 kHz offset <-98 dBc/Hz@1 GHz, 10 kHz offset Amplitude Precision < 0.7 dB < 0.
SSA3000X Spectrum Analyzer Data Sheet Design features Reflection measurement, acquire characteristic curve of the Return Loss EMI filter, Quasi-Peak detector with limit template following CISPR 16 Specifications Specification are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0 ℃ to 50 ℃ temperature, and is warmed up 40 minutes.
SSA3000X Spectrum Analyzer Data Sheet Amplitude Characteristic Amplitude and Level Measurement range DANL to +10 dBm, 100 kHz~1 MHz, preamplifier off DANL to +20 dBm, 1 MHz~3.2 GHz, preamplifier off Reference level -100 dBm to +30 dBm, 1 dB steps Preamplifier 20 dB (nom.), 9 kHz~3.
SSA3000X Spectrum Analyzer Data Sheet Amplitude Characteristic Distortion and Spurious Responses Second harmonic distortion fc≥50 MHz, mixer level -30dBm, attenuation = 0dB, preamp off, 20 ℃ to 30 ℃ -65 dBc Third-order intercept fc≥50 MHz, two -20 dBm tones at input mixer spaced by 100 kHz, attenuation = 0 dB, preamp off, 20 ℃ to 30 ℃ +10dBm 1dB Gain Compression fc≥50 MHz, attenuation = 0 dB, preamp off, 20 ℃ to 30 ℃ >-5 dBm,nom.
SSA3000X Spectrum Analyzer Data Sheet External input and external output Front panel RF input 50 Ω,N-female Front panel TG output 50 Ω,N-female 10 MHz reference output 10 MHz, >0 dBm, 50 Ω, BNC-female 10 MHz reference input 10 MHz, -5dBm~+10dBm, 50 Ω, BNC-female External Trigger input 1 kΩ, 5V TTL , BNC-female Communication Interface USB Host USB-A 2.0 + USB Device USB-B 2.0 LAN LAN (VXI11), 10/100 Base, RJ-45 General Specification Display TFT LCD, 1024×600(waveform area 751×501), 10.
SSA3000X Spectrum Analyzer Data Sheet Ordering Information Product Description Product code Standard configurations Options Optional accessories WWW.SIGLENT.COM SSA3000X Spectrum Analyzer Order Number Spectrum Analyzer, 9 kHz~3.2 GHz SSA3032X Spectrum Analyzer, 9 kHz~2.
SSA3000X Series Spectrum Analyzer About SIGLENT SIGLENT is an international high-tech company, concentrating on R&D, sales, production and services of test &measurement Instruments. SIGLENT began to research and develop the Digital Oscilloscope independently in 2002.