Programming Guide
Table Of Contents
- 1. Programming Overview
- 2. SCPI Overview
- 3. Commands that are Common to All Modes
- 4. Spectrum Analyzer
- 5. Modulation Analyzer
- 6. Programming Examples
SIGLENT
96 SSA3000X Plus Programming Guide
Default
None
Menu
Trace > Copy To
Example
:TRACe:COPY A,B
:TRACe:COPY TRACE1,TRACE2
Command
Format
:TRACe:DEMod:EYE:LENGth <integer>
:TRACe:DEMod:EYE:LENGth?
Instruction
Sets eye length
Gets eye length
Parameter
Type
Integer
Parameter
Range
2 ~ 40
Return
Integer
Default
2
Menu
Trace > Properties
Example
:TRACe:DEMod:EYE:LENGth 4
Command
Format
:TRACe:DEMod:TABLe:FORMat
:TRACe:DEMod:TABLe:FORMat?
Instruction
Display format of Symbol Table data
Parameter
Type
Enumeration
Parameter
Range
BINary| HEXadecimal
Return
Enumeration BIN|HEX
Default
HEX
Menu
Trace > Properties
Example
:TRACe:DEMod:TABLe:FORMat HEX
5.6 Marker Subsection
:TRACe[1]|2|3|4:MARKer[1]|2|3|4:ENABle
:TRACe[1]|2|3|4:MARKer[1]|2|3|4:TYPE
:TRACe[1]|2|3|4:MARKer[1]|2|3|4:X
:TRACe[1]|2|3|4:MARKer[1]|2|3|4:Y?
:TRACe[1]|2|3|4:MARKer[1]|2|3|4:REFerence
:CALCulate[:SELected]:MARKer:COUPle
:CALCulate:MARKer:AOFF